Stryker Tom Tessier Osseous Microtome: Plug for Bone Mill Systems (Parts 01-15400 and 01-15401), Stryker Leiblinger Inc., Kalamazoo, MI; Part 01-15407.
FDA Recall Z-0521-2007 · Posted 2007-02-27
Advertisement
Recalling FirmStryker Leibinger USA
Recall StatusTerminated
Product CodeLXH
Date Initiated2006-12-22
Date Terminated2008-06-04
Product Quantity26
Reason for Recall
Metal shavings come off of the plug as a result of friction between the plug and the cutting cylinder, and may be mixed with the milled bone chips.
Firm Action
Consignees were notified via recall letter dated 12/22/06 which instructed them to quarantine the product and to contact the firm for further instructions.
Root Cause
Other
Distribution Pattern
Nationwide.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.