Medtronic SynchroMed, Model A10
FDA Recall Z-0522-2024 · Posted 2023-12-09
Advertisement
Recalling FirmMedtronic Inc.
Recall StatusCompleted
Product CodeLKK
Date Initiated2020-07-13
Product Quantity2543 units
Reason for Recall
The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM, and Alarm.
Firm Action
Medtronic issued a SAFETY NOTIFICATION to its consignees on 07/13/2020 by letter. The notice explained the problem with the device and requested the software be upgraded to version 1.1.342.
Root Cause
Software Design Change
Distribution Pattern
Foreign Distribution