Ultraview SL Command Module with Masimo Sp02 Option, a multiparameter patient monitor.
FDA Recall Z-0523-2007 · Posted 2007-02-27
Advertisement
Recalling FirmSpacelabs Healthcare, Incorporated
Recall StatusTerminated
Product CodeDSI
Date Initiated2007-01-03
Date Terminated2011-02-15
Product Quantity1347 devices; 1067 domestic and 280 international
Reason for Recall
Potential for modules to reset, alarms returning to default settings, invasive pressure losing its labels and zero calibration and Sp02 alarm limits being frozen. This can result in delay in treatment.
Firm Action
On 1/3/07 the firm issued a letter to their customers advising of the issues and provided guidance to the consignee to enable the Check Setup feature that will provide an audible prompt whenever the monitor resets. The letter advises that consignees will be contacted to schedule an update, once new software has been validated.
Root Cause
Other
Distribution Pattern
Nationwide and worldwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.