Ultraview SL Command Module with Masimo Sp02 Option, a multiparameter patient monitor.

FDA Recall Z-0523-2007 · Posted 2007-02-27

Recall StatusTerminated
Product CodeDSI
Date Initiated2007-01-03
Date Terminated2011-02-15
Product Quantity1347 devices; 1067 domestic and 280 international

Reason for Recall

Potential for modules to reset, alarms returning to default settings, invasive pressure losing its labels and zero calibration and Sp02 alarm limits being frozen. This can result in delay in treatment.

Firm Action

On 1/3/07 the firm issued a letter to their customers advising of the issues and provided guidance to the consignee to enable the Check Setup feature that will provide an audible prompt whenever the monitor resets. The letter advises that consignees will be contacted to schedule an update, once new software has been validated.

Root Cause

Other

Distribution Pattern

Nationwide and worldwide.