Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: OX-M-AEGNGVNTG Usage: Regulator reduces pressure in oxygen-containing cylinder.

FDA Recall Z-0525-2013 · Posted 2012-12-21

Recalling FirmPraxair Inc.
Recall StatusTerminated
Product CodeCAN
Date Initiated2012-11-16
Date Terminated2014-03-24
Product Quantity320,451 units (255,690 US ; 55,884 Canada; 8877 OUS)

Reason for Recall

Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact.

Firm Action

The firm, PRAXAIR, sent a "MEDICAL DEVICE CORRECTION" letter dated November 26, 2012 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to follow the Grab'n Go operating instructions provided in the letter which include: storing the Grab'n Go systems indoors in a safe area; never store a system in an elevated location; handle systems safely and avoid jarring or dropping the system and locate at the following web link: www.praxair.com/GNGinstructions. The customers were also instructed to ensure that all of the Grab'n Go units that were sent by Praxair are sent to Praxair for servicing and refilling after use, in accordance with their normal process. If you have any questions, contact your sales representative at 630-320-4452 or email: [email protected].

Root Cause

Device Design

Distribution Pattern

Worldwide distribution: USA (nationwide) and countries of: Brazil, Canada, China, Columbia, Costa Rica, India, Paraguay and Uruguay.