Philips BV300 Mobile X-Ray System
FDA Recall Z-0526-03 · Posted 2003-02-01
Advertisement
Recalling FirmPhilips Medical Systems
Recall StatusTerminated
Product CodeIZL
Date Initiated2002-09-27
Date Terminated2005-01-12
Product Quantity229 devices
Reason for Recall
The C-bow support can bend and break allowing C-arm to fall from support
Firm Action
The firm issued a letter dated 9/27/02 to their customers. The letter advised of the problem and that a service representative will visit to inspect unit and correct.
Root Cause
Other
Distribution Pattern
The firm distributed to hospitals, x-ray centers and medical centers throughout the U.S.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.