Intubation Flexible Fiberscope, Models: 11301AA1, 11301BN1, 11302BD1, 11302BD2, 11304BC1, 11340BC1, 11301BND1, 11302BDD1, and 11302BDD2. " Transoral intubation " Transnasal intubation " Evaluation of placement of endotracheal and double lumen endotracheal tubes " Provide visual access to the larynx and tracheobronchial tree.
FDA Recall Z-0526-2014 · Posted 2013-12-17
Advertisement
Recalling FirmKarl Storz Endoscopy America Inc
Recall StatusTerminated
Product CodeCAL
Date Initiated2013-11-13
Date Terminated2014-03-27
Product Quantity1516 units
Reason for Recall
KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.
Firm Action
The firm, KARL STORZ, sent a "A SAFETY ALERT-ATTENTION NEEDED" letter dated November 8, 2013 to all their customers on November 13, 2013. The letter describes the product, problem and actions to be taken.The letter informs the customers that the endoscopes incorporated a lumen material that is only approved for use as an external surface material and not included in the STERRAD NX and 100NX FDA-cleared claims for use in flexible endoscope lumens. The customers were instructed to follow the recommended processing methods included in the letter.
Customers with questions about the letter are instructed to contact the undersigned and the Technical Services at (800) 421-0837, ext 5350.
Root Cause
Under Investigation by firm
Distribution Pattern
Nationwide Distribution.