Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600, Model # (UPN) H7879006000, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

FDA Recall Z-0526-2018 · Posted None

Recalling FirmAngiodynamics, Inc.
Recall StatusTerminated
Product CodeNEY
Date Initiated2017-06-05
Date Terminated2018-09-11
Product Quantity14,353 units in total

Reason for Recall

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Firm Action

URGENT VOLUNTARY MEDICAL DEVICE RECALL Letters and Response Forms (dated June 5, 2107) were sent to consignees via FedEx on June 5, 2017. The letter identified the affected product, problem and actions to be taken.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide