CELL-DYN CD 3/4/8 Assay Reagents for Sapphire Hematology Analyzer
FDA Recall Z-0527-06 · Posted 2006-02-16
Advertisement
Recalling FirmAbbott Laboratories
Recall StatusTerminated
Product CodeGKZ
Date Initiated2005-12-16
Date Terminated2006-04-13
Product Quantity871 units
Reason for Recall
T-Cell test results and the respective absolute counts for the CELL-DYN Sapphire instrument using CD3/4/8 reagents can be overestimated by as much as 13%.
Firm Action
The firm has sent notification letters dated Dec. 16, 2005 to all consignees requesting discontinuing of use, destruction of CD3/4/8 reagent if the lab does not plan to run on a CELL-DYN 4000 system.
Root Cause
Other
Distribution Pattern
Nationwide and Canada, Venezuela, Germany, Hong Kong, Japan, Mexico and Singapore.