Alcon Single-Use Curved Jewelers/Iris Bipolar Forceps REF 8065129101 Sterile RX only; Alcon Single-Use Straight/Iris Bipolar Forceps REF 8065129501 Sterile RX only. Usage: Electrosurgical instrument for ophthalmic surgery.

FDA Recall Z-0527-2015 · Posted 2014-12-12

Recalling FirmAlcon Research, Ltd.
Recall StatusTerminated
Product CodeGEI
Date Initiated2014-11-11
Date Terminated2016-02-04
Product Quantity2321

Reason for Recall

The product is sold and labeled as EO sterilized, however the units were not sterilized prior to shipment to customers.

Firm Action

On November 11, 2014, the firm sent letters to their customers notifiying them of the medical device recall. The recalling firm asked their customers to stop using the recalled product, remove it from inventory, return the affected product, and complete a response form.

Root Cause

Process control

Distribution Pattern

Worldwide distribution: US (nationwide) and countries of: Indonesia, South Africa, Switzerland, Brazil, Kazakhstan, and Pakistan.