Optional MDRC-1119 Front Cover Attachment Kit for review display monitors. Product Usage: The product is an optional kit, intended to be attached in front of the LCD panel of a review display (MDRC-1119).

FDA Recall Z-0527-2019 · Posted None

Recalling FirmBarco N.V.
Recall StatusTerminated
Product CodeOWB
Date Initiated2017-03-30
Date Terminated2020-05-04
Product Quantity2,087 units

Reason for Recall

If metal strips are removed and re-applied to another display, there is potential for the metal strips to be deformed resulting in a smaller contact surface with the front cover magnet strips. There is a risk that the protective front cover could come loose and may fall down.

Firm Action

Customers were notified via email on about 03/30/2017. Customers were alerted that a safety related update has been made to the installation manual. A safety notice has been shipped to distributors with the updated installation instructions manual and an acknowledgement letter. Distributors were asked to do the following: a) Inform all your customers who have purchased the Front Cover Kit by sending them the Safety Notice; b) Send the updated installation instructions to all your customers who have purchased the Front Cover Kit; c) Fill out the acknowledgment letter; attach and send in reply to this email.

Root Cause

Device Design

Distribution Pattern

Worldwide - US Nationwide Distribution to CA, IL, GA, TX, and WI. In the countries of Belgium, Brazil, Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Poland, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Taiwan, Turkey.

Official Source

View FDA Recall →