PIVO Blood Collection Device 20G, REF: 202-0005
FDA Recall Z-0527-2022 · Posted None
Advertisement
Recalling FirmVelano Vascular
Recall StatusTerminated
Product CodeJKA
Date Initiated2021-11-10
Date Terminated2023-07-31
Product Quantity123,700
Reason for Recall
Incomplete sealing of blood collection device unit packages may produce a breach in the sterile barrier enclosing the product. This could lead to microbial contamination of the device prior to use. The health effects range from no clinical effect up to bloodstream infection.
Firm Action
On 11/10/21, recall notices were sent to Medical Directors, Risk Managers, Medical Device Safety Officers, Nurse Managers.
1) Locate and destroy all product subject to the recall following your institution s process for destruction. 2) If the affected product was previously used in patient care and no signs of infection are apparent, no additional steps need to be taken. If the affected product was used in patient care and signs of infection are present, follow your institution s established procedures. 3) Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. If you further distributed product, identify customers within your distribution network that purchased the affected product, as defined in this recall notification. Email an excel file listing of all customers to [email protected] so the recalling firm may initiate customer notification. 4) Complete and return the Distributor Response Form. Customers require further assistance are encouraged to call/email: 1-844-823-5433/ [email protected]
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of CT, CO, UT, PA, IL, WA, CA, NJ, MI, FL, NC.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.