Skytron Infinity Series surgical light with dual lightheads, 5 bulbs in 1 - 22' lighthead and 8 bulbs in 1 - 30' diameter lighthead, Manufactured by Dai-ichi Shomei Co., Ltd., Made in Japan; Model IF3022B. (note only the 22'' lighthead is affected.)

FDA Recall Z-0531-2007 · Posted 2007-02-22

Recall StatusTerminated
Product CodeFTD
Date Initiated2007-01-08
Date Terminated2009-11-18
Product Quantity5,000 for all models.

Reason for Recall

If the mounting plug is improperly adjusted, the unit could separate, come apart, and fall from the mount.

Firm Action

The firm initially notified consignees of this problem on 11/23/1999 and again on 11/01/2002 and most lights have already been corrected. Remaining consignees with lights in need of correction were notified of the problem via letter dated 1/8/07 and instructed to contact the distributor to have their lights inspected and modified.

Root Cause

Other

Distribution Pattern

Nationwide and Canada.