Mainline Confirms hCG urine pregnancy test, Mainline Technologies, Inc., Ann Arbor, MI; Catalog # 6008C.

FDA Recall Z-0531-2009 · Posted 2009-01-15

Recall StatusTerminated
Product CodeJHI
Date Initiated2008-08-18
Date Terminated2011-02-07
Product Quantity33919 kits

Reason for Recall

Lack of assurance of safety and efficacy: Unapproved for marketing in the U.S., non-compliance with cGMP regulations and no stability data to support labeled expiration dates.

Firm Action

Consignees were notified via a Medical Device Removal Notice dated 8/18/08 to locate, discontinue use, and return product along with a copy of the letter. The recall was extended to an additional product via letter dated 9/5/08.

Root Cause

PMA

Distribution Pattern

Worldwide Distribution --- including USA and countries of Belgium and Switzerland.