Toshiba Digital Radiography System Model DFP8000D
FDA Recall Z-0532-06 · Posted 2006-02-23
Advertisement
Recalling FirmToshiba American Med Sys Inc
Recall StatusTerminated
Product CodeJAA
Date Initiated2005-11-15
Date Terminated2007-07-12
Product Quantity3
Reason for Recall
In certain conditions the FOV of the lateral plane does not coincide with the X-ray exposed area, and consequently a part of the X-ray field may not be displayed in the magnification mode.
Firm Action
An Advisory Letter will be sent to customers via certified mail. Corrected software will be provided at no charge. Customers can contact their local Toshiba Service person to schedule the software update.
Root Cause
Other
Distribution Pattern
CO, MI, & TX
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.