Toshiba America Medical Systems, Ultimax System MDX-8000A equipped with the EPS-Plus/ADR-1000A/E2D Digital Spot Film Device
FDA Recall Z-0533-06 · Posted 2006-03-29
Advertisement
Recalling FirmToshiba American Med Sys Inc
Recall StatusTerminated
Product CodeLLZ
Date Initiated2006-02-23
Date Terminated2012-02-29
Product Quantity5
Reason for Recall
When the error condition occurs the x-ray exposures will terminate immediately. The digital system (EPS-Plus/ADR-1000A) will immediately stop acquiring images or will continue to acquire blank images. The remote console will lock-up upon release of the hand switch. The operator must reboot the system. This will result in the loss of the images acquired during the contrast portion of the study.
Firm Action
Recall letter to be sent to to customers of affected devices beginning February 23, 2006. Corrected software to be issued by Toshiba America Medical Systems Inc, for all affected devices.
Root Cause
Other
Distribution Pattern
Nationwide to AZ, CA, FL, MT, & PA