COULTER LH 750 Hematology Analyzer.
FDA Recall Z-0536-05 · Posted 2005-02-10
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeGKZ
Date Initiated2004-12-16
Date Terminated2012-04-26
Product Quantity1,257
Reason for Recall
A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation's database are not affected.
Firm Action
Letter 12/20/04 A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation''s database are not affected. If the LH750 instrument displays a pop-up winder titled Database Error (XXXXXX) and text similar to (Microsoft) (ODBC SQL...) As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Do not acknowledge the error. Instead, restart the entire system by completely powering it OFF and then ON. Or, if the LH750 Workstation experiences a change in performance, such as screen display freezes, slow response time, or requires frequent resets to continue processing. As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Verify the accuracy of all patient reports for the samples until service confirms the performance of system.
Root Cause
Other
Distribution Pattern
US and Canada
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.