Microalbumin Multi-Calibrator Set, in vitro diagnostic. Catalog Number: SE-252

FDA Recall Z-0536-2007 · Posted 2007-02-27

Recall StatusTerminated
Product CodeJIX
Date Initiated2006-09-06
Date Terminated2007-05-14
Product Quantity38 kits

Reason for Recall

Microalbumin Multi Calibrator mislabeled label on Level 5 and Level 6, cap label correct

Firm Action

Diagnostic Chemicals Ltd, Canada notified customers by telephone and fax on 9/06/06 and advised to use the cap label to determine concentration of the calibrators. Accounts were asked to confirm and acknowledge receipt of the letter.

Root Cause

Other

Distribution Pattern

Nationwide.