Open, Classified

Platelet Aggregation Kit, Cat. No.5369, For use in platelet aggregation studies, Contains: 2X1 mL ADP Reagent containing adenosine diphosphate, 2X 1 mL Collagen Reagent containing collagen (equine tendon), 2 x 1 mL Epinephrine Reagent containing L-epinephrine, FOR IN-VITRO DIAGNOSTIC USE, Store at 2 to 8 C, 0011724/14(3)

FDA Recall Z-0536-2020 · Posted None

Recall StatusOpen, Classified
Product CodeGHR
Date Initiated2018-05-23
Product Quantity383 Collagen Kits

Reason for Recall

Through an investigation, it has been determined that vials of Collagen packaged within the Collagen Reagent and Platelet Aggregation Kit may contain particulate matter or microbial growth.

Firm Action

On 05/23/2018, the firm's technical services contacted consignees via phone call with a email follow-up of their recall notification letter. Consignees were advised of the following: 1. Prevent any shipment or schedule demonstrations with these lots 2. Remove the lots from available inventory and discard COLLAGEN vials. 3. Verify amount REMOVED FROM STOCK 4. COMPLETE THE FORM AND RETURN TO HELENA LABORATORIES, USA VIA: fax: 409-842-1874 OR E-MAIL TO [email protected]. For questions, contact Jessica W. Jones, MD, MPH - Mgr., Technical Services at: 1 (409) 842-3714 ext. 1177

Root Cause

Unknown/Undetermined by firm

Distribution Pattern

US:TX, WV, OH, TN, GA, TN, NY, ME, MA,LA, AZ OUS: Israel, Korea, Australia, Indonesia, Canada, Thailand, Colombia, Brasil , China , Taiwan