Terminated

IMMULITE 2000 Vitamin B12 Kit, Catalog # L2KVB2, L2KVB6, kit lot 186.

FDA Recall Z-0537-05 · Posted 2005-02-11

Recall StatusTerminated
Product CodeCDD
Date Initiated2005-01-24
Date Terminated2005-07-12
Product Quantity189 kits

Reason for Recall

A low CPS limit was not incorporated in the kit barcode for IMMULITE 2000 Vitamin B12 L2KVB 2, 6 kit lot 186 only.

Firm Action

Domestic customers were notified on January 24, 2005, via telephone and strted receiving a fax-back form and a technical Bulletin on January 25, 2005. Foreign consignees were notified via email and/or Fax.

Root Cause

Other

Distribution Pattern

AL, CA, CO, FL, IL, KS, LA, NY, MA, MD, MI, MS, NC, NE, NM, OH, PA, TN, TX, UT, WA, and Australia, Czech Republic, France, Germany, Netherlands, Portugal, Puerto Rico, Sweden, Turkey, UK, Venzuela