Terminated

Streptex* kit, Ref R30950501 ZL50, Rapid latex test for Streptococcal grouping, IVD. The kit contains reagent groups A, B, C, D, F, G, a control, and an extraction enzyme. The firm name on the label is Remel Europe Ltd., Dartford, England. Streptex is a rapid latex test system for use in the qualitative detection and identification of the Lancefield group of streptococci. Reagents are provided for groups A, B, C, D, F and G covering the majority of clinical isolates, group E streptococci are rarely isolated.

FDA Recall Z-0537-2014 · Posted 2013-12-18

Recalling FirmRemel Inc
Recall StatusTerminated
Product CodeJTO
Date Initiated2013-05-02
Date Terminated2014-03-12
Product Quantity768 kits

Reason for Recall

The product may give weak or false negative test results.

Firm Action

ThermoFisher sent a Medical Device Recall letter dated April 26, 2013, or May 2, 2013, to all affected customers via first class mail. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to notify any personnel who need to be notified of the potential for a false negative test result, review their inventory for affected product lots and discard the affected product. Customers were also instructed to return the attached Medical Device Recall Return Response acknowledging the disposal of the affected product. Customers with questions were instructed to contact the Technical Services Department at 800-255-6730 (US) or 913-888-0939 (International). For questions regarding this recall call 913-888-0939.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution