Terminated

Advanced D-Dimer. Fibrinogen and fibrin split products Catalog numbers OQWM11 and OQEM13.

FDA Recall Z-0538-05 · Posted 2005-02-15

Recalling FirmDade Behring , Inc.
Recall StatusTerminated
Product CodeDAP
Date Initiated2004-12-03
Date Terminated2005-10-18
Product Quantity25381 units

Reason for Recall

decrease in product stability after reconstitution and reduced stability on board certain instrument systems

Firm Action

The recalling firm issued a recall letter to their customers informing them of the problem and the need to use internal quality controls with the analyzers. The letter instructed customer to discard lot # 544545 and reduce the stability for the other lot numbers involved.

Root Cause

Other

Distribution Pattern

The product was shipped nationwide to medical facilities.