Streptex* Latex D, Ref R30950901 ZL54, IVD, Latex reagent for streptococcal group D, 50 tests per box. The firm name on the label is Remel Inc., Lenexa, KS, Manufactured by Remel Europe Ltd., Dartford Kent, UK. Streptex is a rapid latex test system for use in the qualitative detection and identification of the Lancefield group of streptococci
FDA Recall Z-0538-2014 · Posted 2013-12-18
Advertisement
Recalling FirmRemel Inc
Recall StatusTerminated
Product CodeJTO
Date Initiated2013-05-02
Date Terminated2014-03-12
Product Quantity109 kits
Reason for Recall
The product may give weak or false negative test results.
Firm Action
ThermoFisher sent a Medical Device Recall letter dated April 26, 2013, or May 2, 2013, to all affected customers via first class mail. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to notify any personnel who need to be notified of the potential for a false negative test result, review their inventory for affected product lots and discard the affected product. Customers were also instructed to return the attached Medical Device Recall Return Response acknowledging the disposal of the affected product. Customers with questions were instructed to contact the Technical Services Department at 800-255-6730 (US) or 913-888-0939 (International).
For questions regarding this recall call 913-888-0939.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.