Alteon HA Femoral Stem, Press-Fit, Extended Offset, Cementless, HA Coated, Collared, Size 5 All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion. Exactech Alteon HA femoral stems are intended for press-fit fixation.
FDA Recall Z-0538-2018 · Posted 2018-01-18
Advertisement
Recalling FirmExactech, Inc.
Recall StatusTerminated
Product CodeMEH
Date Initiated2017-10-09
Date Terminated2019-08-06
Product Quantity34
Reason for Recall
Wrong device description of the label identifying the implant
Firm Action
Exactech sent an Urgent Medical Device Recall letter dated October 9, 2017, informing them of the potential mislabel. The notice requested consignees: Immediately cease distribution/use of this product; Extend information to accounts in possession of the product; Identify and quarantine any subject devices in their inventory; Complete and return Recall Acknowledgement and Recall Inventory Response Forms. For further questions, please call (800) 392-2832.
Root Cause
Error in labeling
Distribution Pattern
US Distribution to the states of : AL, CA, CO, FL, MD. and NY
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement