Clinical Innovations ROM Plus Collection Kit IFU: ROM Plus Rupture of Membrane Test ROM-5025, ROM-6025

FDA Recall Z-0541-2021 · Posted None

Recall StatusTerminated
Product CodeNQM
Date Initiated2020-08-13
Date Terminated2024-05-24
Product Quantity19450 units/kits

Reason for Recall

Possible false negative or false positive results. The incorrect sample collection swabs (non-scored) were assembled into assembled into the in vitro diagnostic kits. If the user does not break the swab stick correctly, or if scissors or other tools are used to cut the swab handle false negative or false positive results can occur.

Firm Action

On August 17, 2020 through August 19, 2020, the firm sent an "URGENT SAFETY NOTICE" to customers via email informing them that the incorrect swabs (non-scored) were package with their in vitro diagnostic kits and that new instructions and training material will be provided. Customers are instructed to identify any affected products and to complete the response card that is attached to the Customer Notification Letter. Customers are also provided the option of returning the affected products to the Recalling Firm. A company representative will contact and arrange for the return of the product For any questions contact the Recalling Firm at +1-801-268-8200.

Root Cause

Process change control

Distribution Pattern

US: CO, IL, IN, MO, OH, RI, UT, and VA. OUS: None