Roche ISE Internal Standard Gen.2 for use on Roche cobas c systems, Roche Diagnostics, Inc., Indpls., IN; Catalog no. 04522320190. Product is an in vitro diagnostic.
FDA Recall Z-0542-2009 · Posted 2009-01-06
Advertisement
Recalling FirmRoche Diagnostics Corp.
Recall StatusTerminated
Product CodeJIX
Date Initiated2008-09-15
Date Terminated2009-12-01
Product Quantity7909
Reason for Recall
Test results, low: Sodium, potassium and chloride results on the cobas 6000 ISE unit may exhibit a downward drift in controls and patient samples within the recommended calibration interval of 24 hours. For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
Firm Action
Consignees were notified by letter dated 9/15/08, which provided a work-around for them to use until the problem can be permanently resolved. The letter was entitled "URGENT MEDICAL DEVICE CORRECTION." For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
Root Cause
Package design/selection
Distribution Pattern
Nationwide Distribution.