Terumo L Series 1863 type arterial perfusion cannulae, straight open tip, 1/4'' connector, arterial cannula, 10 FR, 17.5 cm (7'') long; Catalog no. L7350.

FDA Recall Z-0543-2007 · Posted 2007-02-27

Recall StatusTerminated
Product CodeDWF
Date Initiated2006-12-07
Date Terminated2007-12-05
Product Quantity50

Reason for Recall

The cannula connector is an incorrect size: It is actually 1/4" by 3/8" instead of the correct 1/4" by 1/4".

Firm Action

Consignee was notified via telephone on 12/7/06 and returned the product.

Root Cause

Other

Distribution Pattern

Pennsylvania.