Varian brand Clinac, Accelerator, Linear, Medical charged-particle radiation therapy system: Model: High Energy Clinacs with one or more of the following options: -- BrainLAB micro MLC, -- 6MV SRS photon beam, -- Fine Beam Isocenter Accuracy, -- Stereotactic motion disable. ----Product is manufactured and distributed by Varian Medical Systems, Palo Alto, CA
FDA Recall Z-0544-2008 · Posted 2008-02-14
Recalling FirmVarian Medical Systems Inc
Recall StatusTerminated
Product CodeIYE
Date Initiated2007-09-10
Date Terminated2010-11-27
Product Quantity418 (total for Recalls Z-0544-0545-2008)
Reason for Recall
Treatment Error: The collimator drive chain may break or slip off of its drive track, allowing the collimator to rotate freely without motor control; if undetected resulting in a treatment with the wrong collimator angle.
Firm Action
On 9/10/07, the firm issued an Urgent Medical Device Correction letter to all its direct consignees, informing them of the device defect and providing instructions on the recall. Short term user instructions were provided and the firm will schedule follow-up visits to make the needed permanent device corrections at each device site.
Root Cause
Labeling design
Distribution Pattern
Worldwide- Canada, Mexico, and the USA