Recall StatusTerminated
Date Initiated2020-11-18
Date Terminated2021-07-22
Product Quantity16 total: 2 units US ; 14 OUS
Reason for Recall
Speakers on IntelliVue X2 and IntelliVue MP2 may fail, identified by the display of a Speaker Malfunct. INOP or by the absence of audible sound, may result in delayed treatment of the patient.
Firm Action
Philips North America issued Customer notifications on 11/18/2020 to US customers, via priority mail. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take:
Philips is asking users to follow the Action to be Taken by Customer/User section of the FSN:
Affected products may continue to be used while awaiting replacement of the speaker, provided that the user routinely checks for the display of the Speaker Malfunct. INOP at power-up.
If you experience this INOP or there is no sound from your X2 / MP2 speaker while it is not connected to another audible alarm annunciation system, remove the device from use and contact your local Philips service representative Philips representative or Philips Customer Support at 1-800-722-9377
Please complete the attached Reply Card and return it to Philips as soon as possible.
Philips will replace affected speaker assemblies free of charge.
A Philips Healthcare representative will contact you to arrange for the replacement of the speaker assembly in the listed affected devices.
Root Cause
Process control
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the state of SD and the countries of China, England, Germany, Hong Kong, Portugal.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K082633INTELLIVUE PATIENT MONITOR, INTELLIVUE XDS, MODELS MP2, MP5, MP20, MP30, MP40, MP50, MP60, MP70, MP80, MP90Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2008-10-03
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.