Datex-Ohmeda S/5 and AS/3 Anesthesia Delivery Units (The S/5 units are a later version of the AS/3 units. The conversion from AS3 to S5 occurred during the summer of 2000. This change was mainly a product name change, with a slight color variation to the shade of white being used externally on the machine. There were no design or feature and function modifications when the name change occurred.)
FDA Recall Z-0545-04 · Posted 2004-07-20
Advertisement
Recalling FirmDatex-Ohmeda Inc
Recall StatusTerminated
Product CodeBSZ
Date Initiated2004-02-03
Date Terminated2005-11-17
Product Quantity1617 units in the United States
Reason for Recall
Anesthetic agents could be delivered at concentrations above the concentration set on the S/5 and AS/3 Anesthesia Delivery Units (ADU) having serial numbers within a specified range.
Firm Action
Recall letters dated February 03, 2004 stated that a firm representative would contact the consignee to schedule a visit to replace software and hardware to correct the problem.
Root Cause
Other
Distribution Pattern
The product was distributed nationwide to hospitals in the Unites States.