Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and CSF, REF CR2003, IVD, UDI: (01)00816387020919

FDA Recall Z-0545-2022 · Posted None

Recall StatusTerminated
Product CodeGMD
Date Initiated2022-01-03
Date Terminated2024-05-15
Product Quantity411 kits

Reason for Recall

The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.

Firm Action

Customers who received product were contacted via letter Urgent: Medical Device Recall, Cryptococcal Antigen Lateral Flow Assay, dated January 3, 2022, beginning January 3, 2022, to locate and immediately destroy their affected devices. A recall notification and response form was attached to the email that is to be completed by the customers and returned to the firm.

Root Cause

Material/Component Contamination

Distribution Pattern

US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑