Continuous Peripheral Nerve Block Catheter Kit and Set Continuous Peripheral Nerve Block Catheter Kit and Set permits placement of catheters next to nerves and nerve plexuses for continuous nerve block anesthesia or analgesia techniques for periods not exceeding 72 hours.

FDA Recall Z-0546-2014 · Posted 2013-12-19

Recall StatusTerminated
Product CodeCAZ
Date Initiated2013-10-09
Date Terminated2015-09-21
Product Quantity121

Reason for Recall

A labeling inconsistency was discovered in which the lidstock states that the kit contains a stimulator cable with 1.5mm socket, however, the kit contains a stimulator cable with 1.5mm pin.

Firm Action

The firm, Arrow International, Inc., sent an "Urgent Medical Device Recall Notification" letter dated October 14, 2013 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use and quarantine any of listed affected products/lots you have in stock; return affected products, customer service representative will contact you with RGA# and instructions for the return of the product, and complete enclosed Recall Acknowledgement form and fax to 1-800-343-2935, Attn: Customer Service, even if you have not affected product. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-800-343-2935.

Root Cause

Labeling mix-ups

Distribution Pattern

Worldwide distribution: US (nationwide) including states of: MA, LA, CA, GA, KS, TX, PA, NY, MO, WI, NJ, MS, CT, ID, AL, CO, and PR; and Singapore.