Randox Rx Series Mg - Product Usage: For the quantitative in vitro determination of Magnesium in serum, plasma and urine Catalogue Number: MG8326
FDA Recall Z-0547-2021 · Posted None
Advertisement
Recalling FirmRandox Laboratories Ltd.
Recall StatusTerminated
Product CodeJGJ
Date Initiated2020-11-13
Date Terminated2021-04-01
Product Quantity4 kits
Reason for Recall
Failing to meet the linearity performance claims
quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine
Firm Action
Randox issued Urgent Medical Device Correction letter via e-mail to US Customers on 13 November 2020 stating reason for recall, health risk and action to take:
Discontinue use of and discard any of the above immediately.
" Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory.
" Review results generated with the affected batches in line with the clinical profile of the patient.
" Discuss the contents of this notice with your Medical Director.
" Complete and return the response form 12187-QA to technical. [email protected] within five working days.
Questions or concerns, please contact Randox Technical Services.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the states of GA, NJ.