INTEGRA SURFIX ALPHA Screwdriver Torx 10, REF 219 135 ND
FDA Recall Z-0548-2022 · Posted None
Advertisement
Recalling FirmSmith & Nephew, Inc.
Recall StatusOpen, Classified
Product CodeHRS
Date Initiated2022-01-04
Product Quantity8 units (US)
Reason for Recall
The affected screwdrivers are out of specification and may not mate with the Surfix Alpha Lock screw
Firm Action
The firm initiated the recall by letter and email on 01/04/2022. The letter requested that the consignee examine their inventory, quarantine all affected units, and return the completed response form. Entities who have further distributed the product were directed to notify their customers. The firm is seeking return of the product.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of CA and OH.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.