Clark Biocompatible Hemoperfusion Cartridge, Sorbent: Heparinized polymer over acitivated carbon, 175 mL, Clark Research and Development, Inc. Folsom, LA 70437
FDA Recall Z-0552-2007 · Posted 2007-02-28
Advertisement
Recalling FirmClark Research and Development Inc
Recall StatusTerminated
Product CodeFLD
Date Initiated2007-01-19
Date Terminated2007-07-17
Product Quantityunknown
Reason for Recall
Lack of sterility assurance based on lack of sterility validation for labeled sterilization directions
Firm Action
The firm notified its consignees of the problem and the recall via letter (priority mail with delivery confirmation)dated 01/19/2007 beginning on 01/19/2007. This letter requested that the consignee cease the use of this product and contact the recalling firm for return instructions. Follow-up with non-reponders will be made by facsimile.
Root Cause
Other
Distribution Pattern
Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.