Biomet 20mm Compress Device Short Anchor Plug With Drill Set / With Pin Inserter Ion Implant Item number: 178562 Product Usage: 1. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. 2. Tumor resections. 3. Revision of previously failed total joint arthroplasty. 4. Trauma. The Compress Segmental Femoral Replacement System (Short Spindle and Anchor Plug) components are intended for uncemented use only.

FDA Recall Z-0553-2019 · Posted None

Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeJDI
Date Initiated2018-10-24
Date Terminated2020-04-14
Product Quantity6 units

Reason for Recall

Potential metal burrs in the holes of the device preventing the appropriate drill and transverse pins from being inserted

Firm Action

Zimmer Biomet initiated recall by letter on 10/24/18 to distributors and hospitals advising of the problem and action to be taken-locate quarantine and return product. Complete Inventory Return Certification Form. Questions or concerns call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday.

Root Cause

Employee error

Distribution Pattern

US Nationwide Distribution in the states of IN, DC, MD

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.