Medline Custom Sterile Surgical Packs which contain a Tyvek Breather Pouch, part number 32901, lot numbers 44134-01 and 43589-01; the breather pouch measures 25.25 inches x 30.75 inches, 5 mil thick; Medline Industries, Inc., Mundelein, IL 60060;
FDA Recall Z-0554-05 · Posted 2005-03-11
Advertisement
Recalling FirmMedline Industries, Inc.
Recall StatusTerminated
Product CodeLRO
Date Initiated2005-03-04
Date Terminated2006-05-25
Product Quantity31,497 cases
Reason for Recall
There is a potential for open side seals post-sterilization of the Tyvek breather pouches used in the custom sterile packs.
Firm Action
Field Correction letters dated 3/4/05 were sent to thedirect account hospitals on the same date to the attention of the manager of surgery, informing them of the potential indadequate side seal of the Tyvek breather pouches in the packs which may cause a compromise in the sterility of the product. The accounts were provided a list of the affected pack sent to the account and were requested to inspect any stocks of the affected product in stock, using the enclosed diagrams as an aid in the inspection for the defective side seals. Any defective product found should be returned to Medline. Any questions were directed to Lara Simmons, QA/RA Director at 1-800-879-2781, ext. 6072.
Root Cause
Other
Distribution Pattern
Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.