Zimmer / Abbott Spine Sequoia dorsal height & revision tool, non-sterile; REF 3367-1. Surgical instrument for pedicle screw system.
FDA Recall Z-0554-2010 · Posted 2009-12-22
Advertisement
Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeNKB
Date Initiated2009-08-20
Date Terminated2011-04-13
Product Quantity215
Reason for Recall
The tip of the instrument may fracture during use.
Firm Action
Zimmer sent a letter dated 8/17/09 to distributors and consignees on or about 8/20/09. Users were notified of the potential problem via the letter and advised that the instruments will be removed and replaced when a new version of the instrument becomes available (around 10/09).
Root Cause
Device Design
Distribution Pattern
Nationwide, Belgium and New Zealand.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.