Terminated

Merge Hemo with software versions 10.0 LA. Merge Hemo (formerly named HeartSuite Hemodynamics) monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure.

FDA Recall Z-0554-2016 · Posted None

Recall StatusTerminated
Product CodeDQK
Date Initiated2015-10-26
Date Terminated2016-07-22
Product Quantity2

Reason for Recall

The firm, Merge Healthcare, sent out Merge HEMO V10.0 & Merge CARDIO V10.1 software to 3 customers before the product was completely validated in house.

Firm Action

Consignees were either contacted via phone 9/2015 or in person 7/31/2015. The customer communications involved the need to upgrade the consignees software. Merge Healthcare did not script the conversations because they opted to hold one-on-one management level discussions to carefully explain the importance of moving to the Generally Available (GA) release.

Root Cause

Process control

Distribution Pattern

Distributed in the states of IL, NC, and VT.