The Triathlon Posteriorly Stabilized Femoral Component.
FDA Recall Z-0556-05 · Posted 2005-02-24
Advertisement
Recalling FirmStryker Howmedica Osteonics Corp.
Recall StatusTerminated
Product CodeNPJ
Date Initiated2004-10-27
Date Terminated2005-05-31
Product Quantity386 UNITS
Reason for Recall
The fatigue testing of the Triathlon PS Femoral Component does not consistently meet the requirements of the finite element analysis predicted load as delineated in the 510(k) submittal.
Firm Action
Notification letters and Product Accountability Froms have been sent via Fedex on 10/27/2004 with return receipt.
Root Cause
Other
Distribution Pattern
The product is distributed to its branches/agencies and to hospitals nationwide.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.