The ETI-Mumps IgG Enzyme Immunoassay Test Kit is sold with assay software version 1.0. that is designed to work exclusively with the ETI-Max 3000 automated platform (version 1.51.1).
FDA Recall Z-0557-06 · Posted 2006-02-23
Advertisement
Recalling FirmDiasorin Inc.
Recall StatusTerminated
Product CodeLJY
Date Initiated2005-08-08
Date Terminated2006-07-22
Product Quantity22
Reason for Recall
Potential for false negative results.
A software anomaly may occur when performing the ETI-Mumps IgG assay on the ETI-Max 3000 automated platform. The error occurs when running a combined plate on the ETI-Max 3000, with the ETI-Mumps IgG assay as the second assay on a combined plate with ETI-Measles IgG and/or ETI-VZV IgG.
Firm Action
Consignees were contacted by telephone and then the firm sent a fax titled 'Diasorin Customer Notification' dated 8/8/05 that documented the information as provided by telephone.
Root Cause
Other
Distribution Pattern
Nationwide. AK, CA, CT, DC, KS, MA, NY, NJ, PA, SD, TN, TX.