OEC FluoroTrak Plus 9800 Fluoroscopic X-ray System,

FDA Recall Z-0557-2007 · Posted 2007-03-22

Recall StatusTerminated
Product CodeJAA
Date Initiated2007-02-09
Date Terminated2007-06-26
Product Quantity11 units of all models

Reason for Recall

During routine service, a cover may have been installed without the required proper lead shielding.

Firm Action

Consignees were notified by telephone on 02/09/2007 and told to discontinue use of the units until they could be inspected by GE personnel. GE personnel visited all 8 sites on 02/12/2007.

Root Cause

Other

Distribution Pattern

Nationwide to AL, AZ, IA, OH, TX, VA, and WV.