Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 20cm, Material CA20L1
FDA Recall Z-0558-2018 · Posted None
Advertisement
Reason for Recall
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
Firm Action
All consignees were notified via Federal Express or certified mail on August 7, 2017, and the letter informs customers of the possible trocar tip disengagement and the actions they should take. Customers are requested to segregate and return affected product and acknowledge receipt and understanding of the Urgent Medical Device Recall Notice.
Root Cause
Device Design
Distribution Pattern
Nationwide including PR, Canada, China
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.