MANI Infusion Cannula 23G S - Product Usage: for use in injecting perfusate or aspirating out intraocular materials during ophthalmic surgery Ref: MTC 23S
FDA Recall Z-0558-2021 · Posted None
Advertisement
Recalling FirmMani, Inc. - Kiyohara Facility
Recall StatusTerminated
Product CodeNGY
Date Initiated2020-10-27
Date Terminated2021-08-11
Product Quantity60 units (US)
Reason for Recall
Dimensions of the Luer connector parts out of specification, may cause leakage of fluid and unintended disconnection during use
Firm Action
MANI Inc. issued Voluntary Removal letter issued on 10/27/20 via email to the US Distributor stating reason for removal, health risk and action to take: request distributors to return the devices. If there are MANI TROCAR KIT and MANI INFUSION CANNULA in your stock and/or your customer's stock: Please write the returning Q'ty. in the list ,"Returned" If MANI TROCAR KIT and MANI INFUSION CANNULA are already used up : Please write the used Q'ty. in the list , "Used.
Please send this sheet and relevant items if you have to our address below.
Any costs/fees related to this return shipment should be charged to MANI FedEx account.
MANI, INC.
8-3, Kiyohara Industrial Park, Utsunomiya, Tochigi, 321-3231, Japan
Root Cause
Unknown/Undetermined by firm
Distribution Pattern
US Nationwide distribution in the state of MO.