OEC 9800MD Motorized C-arm System, Fluoroscopic X-ray System,
FDA Recall Z-0559-2007 · Posted 2007-03-22
Advertisement
Recalling FirmGE OEC Medical Systems, Inc
Recall StatusTerminated
Product CodeJAA
Date Initiated2007-02-09
Date Terminated2007-06-26
Product Quantity11 units of all models
Reason for Recall
During routine service, a cover may have been installed without the required proper lead shielding.
Firm Action
Consignees were notified by telephone on 02/09/2007 and told to discontinue use of the units until they could be inspected by GE personnel. GE personnel visited all 8 sites on 02/12/2007.
Root Cause
Other
Distribution Pattern
Nationwide to AL, AZ, IA, OH, TX, VA, and WV.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.