Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging.

FDA Recall Z-0561-2010 · Posted 2009-12-23

Recall StatusTerminated
Product CodeJAK
Date Initiated2008-09-17
Date Terminated2009-12-23
Product Quantity82

Reason for Recall

The system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the x-ray and gantry motions will both resume.

Firm Action

A letter dated 9/17/08 was sent to each account telling them that the PLC would be upgraded. The firm visited each account and made corrections.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution