Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Evolut (TM) PRO+ System. With the following model numbers: a. EVPROPLUS-23US, Labeled as VLV EVPROPLUS-23US COMM JPN, VLV EVPROPLUS-23 COMM US; b. EVPROPLUS-26US, Labeled as VLV EVPROPLUS-26US COMM JPN, VLV EVPROPLUS-26 COMM US; c. EVPROPLUS-29US, Labeled as VLV EVPROPLUS-29US COMM JPN, VLV EVPROPLUS-29 COMM US; d. EVPROPLUS-34US, Labeled as VLV EVPROPLUS-34 COMM US, VLV EVPROPLUS-34 COMM US SA.

FDA Recall Z-0561-2021 · Posted None

Recall StatusTerminated
Product CodeNPT
Date Initiated2020-10-23
Date Terminated2022-11-22
Product Quantity19,961 units

Reason for Recall

Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).

Firm Action

On 10/23/2020, the firm sent an "Urgent Medical Device Correction" Notification Letter to customers via UPS 2-day delivery informing them that the Recalling Firm has received complaint of valve leaflet damage occurring following a post-implant balloon dilatation (PID) resulting in moderate or severe aortic insufficiency which were detected acutely or during follow up. The Recalling Firm has identified the potential cause of the damage which is due to over-expansion of the narrowest portion (waist) of the Transcatheter Aortic Valves (TAVs). The Recalling Firm is also stating that depending on the choice of balloon, the physicians must consider two factors that may lead to over-expansion of the waist of the TAV: 1.The pressure the balloon is inflated to when performing PID 2.Balloon size used for PID The Recalling Firm is providing detailed guidance on considering the above two facts and is providing in an Appendix (Appendix A) with the Customer Notification Letter. In addition, the Appendix A is providing updated precautionary instructions related to PID. The Recalling Firm is also informing its customers that for patients who have been, or will be, treated with the affected product, should continue to be managed according to their standard patient management protocols. Physicians are also asked to: - Review the updated instructions provided in Appendix A. - Share this information with other physicians in your facility who use the affected product. - After review of this information, complete the enclosed Physician Confirmation Certificate and email to [email protected]. Any Questions can be directed to the Recalling Firm's Field Representatives or Lifeline Technical Services at 1-877-526-7890, Monday-Friday, 8am-5pm Central Time.

Root Cause

Labeling design

Distribution Pattern

Worldwide distribution - U.S. Nationwide distribution and the countries of Panama, Denmark, Austria, Germany, Finland, Sweden, Colombia, India, Mexico, Brazil, Spain, United Kingdom, Australia, Ireland, Italy, Argentina, Japan, Poland, Russian Federation, Hungary, Kazakhstan, Luxembourg, Portugal, Norway, Egypt, Saudi Arabia.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑