LIFEPAK 500 AED.

FDA Recall Z-0562-05 · Posted 2005-03-10

Recall StatusTerminated
Product CodeMKJ
Date Initiated2005-02-03
Date Terminated2006-08-21
Product Quantity2,349

Reason for Recall

Potential to not detect patients at the low end of the patient impedance range.

Firm Action

On 2/3/05 the firm issued a 'Medical Device Correction' letter dated January 2005 to their customers. The letter advised customers of the issue and stated a representative from the firm would visit each facility to update affected unit. On 2/25/05 a nationwide press release was issued. On 3/1/05 the firm began telephoning each customer, this was followed by another letter titled MEDICAL DEVICE RECALL. The letter advises of the issue and recommends defibrillators remain in service when no alternative is available until the unit can be corrected. Each unit will be replaced or corrected. On 4/22/05 the firm issued another letter, via certified mail return receipt, expanding their recall. Letter was issued to all customers affected by the expansion of the recall. On 4/22/05 the firm issued a press release advising of the expansion of the recall. The firm's web site also provides a listing of affected units.

Root Cause

Device Design

Distribution Pattern

Devices were distributed to 750 domestic consignees dn 16 international consignees.

Official Source

View FDA Recall →