Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.
FDA Recall Z-0563-06 · Posted 2006-02-28
Advertisement
Recalling FirmCygnus LLC
Recall StatusTerminated
Product CodeJQC
Date Initiated2005-02-04
Date Terminated2007-02-01
Product Quantity20 units
Reason for Recall
The units were exhibiting tolerance stacking in the electronic circuit timer that controls the electric motor that spins the centrifuge and applies the centrifuge brake.
Firm Action
A recall notificaiton was sent to the firm''s customers by e-mail on 2/14/2005. Distributor conducted sub-recall.
Root Cause
Other
Distribution Pattern
All product was shipped domestically to the firm''s distributor, Bock Scientific, 1170 Lincoln Avenue, Unit 11, Holbrook, NY 11741
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.