Terminated

Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.

FDA Recall Z-0563-06 · Posted 2006-02-28

Recalling FirmCygnus LLC
Recall StatusTerminated
Product CodeJQC
Date Initiated2005-02-04
Date Terminated2007-02-01
Product Quantity20 units

Reason for Recall

The units were exhibiting tolerance stacking in the electronic circuit timer that controls the electric motor that spins the centrifuge and applies the centrifuge brake.

Firm Action

A recall notificaiton was sent to the firm''s customers by e-mail on 2/14/2005. Distributor conducted sub-recall.

Root Cause

Other

Distribution Pattern

All product was shipped domestically to the firm''s distributor, Bock Scientific, 1170 Lincoln Avenue, Unit 11, Holbrook, NY 11741