Terminated

Centrifuge Model CF-6 with lid lock. (Block Model Number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.

FDA Recall Z-0564-06 · Posted 2006-02-28

Recalling FirmCygnus LLC
Recall StatusTerminated
Product CodeJQC
Date Initiated2005-09-08
Date Terminated2007-04-17
Product Quantity110 units

Reason for Recall

When device was packaged, too much expanding foam was added causing the lid to deform and causing the lid locking mechanism to be out of alignment.

Firm Action

A recall notice was sent to Bloch Scientific on September 8, 2005. Bloch further distributed the product to MTF, Cascade and will conduct a sub-recall

Root Cause

Other

Distribution Pattern

All product was shipped domestically to the firm''s distributor: Block Scientific, 1170 Lincoln Ave., Unit 11, Holbrook, NY 11741