Centrifuge Model CF-6 with lid lock. (Block Model Number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.
FDA Recall Z-0564-06 · Posted 2006-02-28
Advertisement
Recalling FirmCygnus LLC
Recall StatusTerminated
Product CodeJQC
Date Initiated2005-09-08
Date Terminated2007-04-17
Product Quantity110 units
Reason for Recall
When device was packaged, too much expanding foam was added causing the lid to deform and causing the lid locking mechanism to be out of alignment.
Firm Action
A recall notice was sent to Bloch Scientific on September 8, 2005. Bloch further distributed the product to MTF, Cascade and will conduct a sub-recall
Root Cause
Other
Distribution Pattern
All product was shipped domestically to the firm''s distributor: Block Scientific, 1170 Lincoln Ave., Unit 11, Holbrook, NY 11741
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.