Terminated

Zimmer Orthopaedic Surgical Products, Disposable Cuff, Single Port, Double Bladder Cuff, Sterile, Zimmer, Dover, OH USA. The cuff is a sterile, single use product equipped with an integral fill line. The product is packed 10 cuffs per unit.

FDA Recall Z-0565-05 · Posted 2005-02-25

Recall StatusTerminated
Product CodeKCY
Date Initiated2005-02-11
Date Terminated2006-04-07
Product QuantityTotal 430 cuffs

Reason for Recall

The port/tube bond in the affected cuffs may separate during surgery causing the pressure in the cuff to release.

Firm Action

Consignees were notified by letter on 02/11/2005.

Root Cause

Other

Distribution Pattern

Nationwide